Quantifies ECG quality per channel by the agreement between two independent QRS detectors. A clean signal makes both detectors agree on nearly every beat (bSQI near 1); noise makes them disagree (bSQI drops).

Usage

ecgBeatSQI(
  x,
  detector_a = "pan_tompkins",
  detector_b = "elgendi",
  match_window_ms = 150,
  threshold_factor = 0.6,
  refractory_ms = 200,
  beta = 0.08,
  assay_name = NULL
)

Arguments

x

A PhysioExperiment object with ECG data.

detector_a, detector_b

Detector methods to compare (see ecgDetectRpeaks); defaults "pan_tompkins" and "elgendi".

match_window_ms

Two beats match when within this window (default 150).

threshold_factor, refractory_ms, beta

Passed to the detectors.

assay_name

Assay to use (default: the default assay).

Value

A data.frame with one row per channel: n_a, n_b (beat counts), n_matched, and bSQI = matched / (n_a + n_b - matched) (Jaccard agreement in [0, 1]; NA for a silent channel).

References

Li, Q., Mark, R.G. & Clifford, G.D. (2008). Robust heart rate estimation from multiple asynchronous noisy sources. Physiological Measurement, 29(1), 15-32.

Examples

pe <- make_ecg(n_time = 5000, sr = 500, heart_rate = 72)
ecgBeatSQI(pe)
#>   channel n_a n_b n_matched bSQI
#> 1       1  11  11        11    1