Fits the Konno-Mead two-degree-of-freedom volume model from ribcage and abdominal belt signals. Least-squares calibration uses a reference volume trace; qualitative diagnostic calibration (QDC) equalizes the variability of the two belt contributions.
Arguments
- rc
Ribcage signal as a numeric vector, or a
PhysioExperiment.- ab
Abdominal numeric vector, or a channel index/label when
rcis aPhysioExperiment.- reference
Optional reference volume signal, required for
"lsq".- sr
Sampling rate in Hz for numeric signals. A
PhysioExperimentsupplies its own sampling rate.- method
Calibration method,
"qdc"or"lsq".- band
Two-element respiratory band in Hz.
- window
Optional inclusive calibration sample range
c(lo, hi).- fit_intercept
Logical; include an intercept in least squares.
- known_volume
Optional known inspiratory volume used for absolute scaling.
Value
A rip_calibration object containing coefficients, calibration
diagnostics, and the calibrated volume-proportional signal.
References
Konno K, Mead J (1967). Measurement of the separate volume changes of rib cage and abdomen during breathing. Journal of Applied Physiology, 22:407-422.